We work through every interrogation using the PBL-STOP framework — a quick mental checklist so nothing gets missed. The ILR is a monitor, not a therapeutic device, so we use an abbreviated version: P, B, L (sensing only), S (the main event), and P. T and O are explicitly skipped.
Device overview
An implantable loop recorder is a small subcutaneous monitor placed in the left parasternal area. It records a single-channel ECG continuously and stores episodes triggered automatically by algorithm or by the patient with a handheld activator or smartphone app. Most ILR follow-up is remote, with in-office visits reserved for specific issues — sensing problems, episode review for borderline findings, or end-of-indication decisions. Indications are narrow but important: unexplained syncope, palpitations without documented arrhythmia, cryptogenic stroke (AF surveillance), and post-ablation AF monitoring.
PBL-STOP walkthrough (abbreviated)
P — Presentation / Pre-interrogation review
- Manufacturer, model, serial, implant date
- Indication: syncope, palpitations, cryptogenic stroke, post-ablation AF surveillance
- Symptom diary entries since last review
- Episodes already adjudicated at prior transmissions
- Current rhythm-relevant medications (rate-control, antiarrhythmics, anticoagulation status)
- Outstanding clinical questions: are we still looking for the same arrhythmia, or has the clinical situation evolved?
A 30-second chart pass before opening the episode list keeps the review focused — you want to know what question the ILR is answering before you start clicking through strips.
B — Battery / longevity
- Voltage and estimated remaining longevity
- Expected total service typically 3–4.5 years depending on platform and episode volume
- Plan ahead:
- >12 months remaining: continue monitoring per indication
- 6–12 months: confirm whether the clinical indication still exists
- <6 months: plan explant vs replacement based on whether the diagnostic question is unresolved
ILR replacement is uncommon — most devices are pulled at diagnosis or end-of-indication. Battery alone rarely drives a replace-vs-explant decision; the clinical question does.
L — Lead measurements (sensing margin only)
The ILR has no pacing leads — there is no impedance, no capture threshold. The L section reduces to sensing.
- R-wave amplitude: should be >0.3 mV
- Trend it across visits — declining sensing is the leading cause of new false positives
- Sustained R-wave <0.2 mV will degrade every algorithm downstream
- Body position effects — some platforms show sensing in supine vs upright
- Note any signal saturation, baseline drift, or noise artifact patterns in the strip
- If sensing has dropped meaningfully, consider:
- Device migration
- Scar maturation around the electrodes
- Patient body habitus change
- Need to re-position or reprogram sensing gain
S — Stored events / Sensed rhythm
This is the main event of an ILR check — everything else exists to support this section.
Triage order
- Patient-activated symptomatic episodes first — these are tied to symptoms and have the highest diagnostic value
- Then auto-detected episodes by clinical severity:
- Asystole / long pause
- AF
- Sustained tachycardia
- Bradycardia
- Non-sustained tachycardia
EGM review for every flagged episode
- AF claims: irregular RR with no organized P-waves vs noise vs SVT vs frequent PACs
- Pause claims: true asystole vs R-wave undersensing vs lead artifact vs T-wave/P-wave dropout
- Tachycardia claims: sinus vs SVT vs VT — width matters, onset matters
- Bradycardia claims: true sinus brady vs sensed pause from undersensing
Symptom correlation
- Cross-reference each device episode with the patient diary
- Symptomatic episode + documented arrhythmia = diagnostic
- Symptomatic episode + sinus rhythm = also diagnostic (rules out arrhythmic cause)
- Asymptomatic AF in a cryptogenic stroke patient = anticoagulation discussion
Memory management
- Clear reviewed episodes to free storage
- High event volume between transmissions can overflow memory and lose data
- If overflow is happening, tighten the detection thresholds or shorten the transmission interval
Detection threshold adjustments
- Tighten if false positives are overwhelming the workflow
- Loosen if the clinical question demands maximum sensitivity (e.g., active cryptogenic stroke workup)
T — Therapy
Not applicable — the ILR does not deliver therapy or pace.
O — Optimization
Not applicable — the ILR does not deliver therapy or pace.
P — Plan / Programming changes
- Document every threshold change with prior value and rationale
- Save EGM strips of any clinically relevant episode to the EMR
- Reset counters and clear reviewed episodes
- Patient education refresh: activator use, what to do if symptomatic, when to call
- Set next remote transmission and next in-office visit
- Escalate to the EP attending for:
- Asystole >6 seconds, regardless of symptoms
- AF with stroke risk and no anticoagulation
- Any wide-complex tachycardia
- Syncope with a corresponding device-recorded arrhythmia
- Worsening symptoms with no device correlation — may need a different monitoring strategy
- End-of-indication explant decisions
A note on volume
The biggest workflow challenge with ILRs is volume. A single patient with a borderline AF algorithm can generate dozens of false positives per week. Triage policies — who reviews what, what gets escalated, and how often we revisit detection thresholds — keep the device clinic functional. A trained ILR reader who knows the difference between noise, pseudo-AF from PACs, and true AF saves the practice an enormous number of hours.