We work through every interrogation using the PBL-STOP framework — a quick mental checklist so nothing gets missed. On an ICD the T section carries the most weight, because the device’s job is therapy decisions and every one of those decisions deserves an EGM review.
Device overview
An ICD does everything a pacemaker does, plus it monitors for ventricular tachycardia and fibrillation and delivers ATP or shocks when criteria are met. Every clinic visit answers two questions: is the device working, and did the device make any decisions since the last visit — and were they correct?
PBL-STOP walkthrough
P — Presentation / Pre-interrogation review
- Manufacturer, model, serial, implant date
- Indication: primary prevention (low EF, ischemic vs non-ischemic) vs secondary prevention (prior VT/VF arrest)
- Programmed zones at last visit (VT monitor / VT / VF cutoffs)
- Shock history since implant — count, last shock date
- Remote alerts since last in-person check (high-voltage impedance, RV lead noise, AHRE)
- Current antiarrhythmics and beta-blocker dosing
- Patient-reported symptoms: palpitations, near-syncope, perceived shocks
B — Battery / longevity
- Voltage compared against model-specific RRT/ERI threshold
- Percent remaining, projected months of service
- BOL / MOL / ERI / EOS status
- Drivers of accelerated depletion specific to ICDs:
- Frequent shock deliveries (each charge is expensive)
- Frequent capacitor reforming
- High RV pacing burden — bad for the patient and the battery
- High pacing outputs in CRT-D configurations
Devices with multiple recent shocks may have a markedly shorter projected longevity than the prior visit indicated — always recompute after a shock cluster.
L — Lead measurements
Per lead, with trends:
- RA and RV pacing impedance: 400–1200 ohms
- RV shock (high-voltage) impedance: 30–80 ohms
- Change >20 ohms suggests a coil or HV conductor issue
- Capture thresholds at 0.5 ms
- Sensing amplitudes:
- R-wave typically >5 mV
- Drop >50% from baseline warrants EGM review for undersensing
- Sudden swing in any of these is more important than the absolute number
S — Stored events / Sensed rhythm
- Episode log organized by zone and chronology
- For each episode: stored EGM, intervals, classification
- AT/AF burden — episode count, total duration, fastest atrial rate
- AHRE >6 minutes triggers an anticoagulation discussion
- NSVT in the monitor zone — frequency, fastest rate
- Pacing percentages and rate histograms
- Patient-triggered transmissions and surrounding rhythm
T — Therapy
This is the heart of an ICD interrogation. Slow down here.
Detection zones — confirm and rationalize
- VT monitor zone: counts only, no therapy
- VT zone: cutoff typically 170–188 bpm, ATP then shocks
- VF zone: cutoff typically 188–220 bpm, shocks (with ATP during charging on many platforms)
- Long detection (30/40 intervals or ~12 seconds) reduces both inappropriate shocks and avoidable shocks for self-terminating runs
SVT discriminators — confirm enabled
- Onset (sudden vs gradual)
- Stability (regular vs irregular RR)
- Morphology (template match against stored sinus QRS)
- A:V relationship in dual-chamber/CRT-D devices
Episode-by-episode appropriateness review
For every therapy delivered since the last visit, walk the EGM:
- Was the rhythm truly VT or VF?
- Were the intervals consistent with the classification?
- For ATP: did it terminate, was it ineffective, or did it accelerate to VF?
- For shock: did it convert? On the first shock or after redetect?
- Was sensing clean — no T-wave oversensing, no double-counting, no lead noise masquerading as VF?
- For inappropriate shocks, classify the cause:
- AF with rapid conduction
- Sinus tachycardia tracking through discriminators
- SVT (AVNRT, AVRT, AT)
- Lead noise or fracture
- T-wave oversensing
- EMI
Post-shock pacing
- Confirm a post-shock pacing rate (often 90 bpm) is programmed to bridge bradyasystole during stunned myocardium
- Confirm post-shock outputs are set with adequate safety margin — captures matter more than battery here
- Duration of post-shock pacing (often 30 seconds to several minutes)
Manual testing as indicated
- Capture threshold at 0.5 ms
- Sensing test in VVI
- Morphology template re-acquisition if the QRS has changed since implant
- DFT testing is rarely repeated in clinic — flag to the attending if indicated
O — Optimization
- Minimize RV pacing in primary-prevention ICDs — this device is for shocks, not pacing
- AV delay set to favor intrinsic conduction
- Mode switch sensitivity calibrated to real AT, not far-field oversensing
- Rate response off unless chronotropic incompetence is documented
- Detection duration lengthened where appropriate to reduce avoidable shocks
- SVT discriminators tightened or loosened based on the appropriateness review
P — Plan / Programming changes
- Document every change with the prior value and the rationale
- Re-interrogate after the change
- Reset counters per institutional policy
- Schedule next remote transmission and in-office visit
- Patient education after any shock — appropriate or not. Confirm action plan: when to call, when to go to the ED, what one shock vs multiple shocks means.
- Escalate to the EP attending for:
- Any shock (same-day or next-day call)
- Inappropriate detection from oversensing
- HV impedance alert or lead fracture pattern
- VT cluster (≥3 episodes in 24 hours) — VT storm protocol
- Failed shock — vector, position, energy review
- Phantom shock with no device-recorded therapy — exclude noise first